Skip to content

 

StratX® CT Parameters

The following CT scan radiographic parameters have been optimised in order to provide appropriate input to generate a StratX Lung Analysis Report.

In order to produce quality output parameters displayed in the StratX Report, it is recommended to adhere as closely as possible to the ideal parameters. If a scan can’t be obtained with the ideal parameters, we have specified acceptable scan parameters below. Significant deviation from these parameters may result in reduced accuracy or an inability to analyse the CT scan.

For more specific information related to scan parameters or if your CT scanner manufacturer is not listed below, please contact us directly for support.

StratX Report Key Upload Parameters

  1. Ensure all files are in standard .DICOM format.
  2. Upload axial scan series only to reduce upload time and amount of data transferred (reduce data corruption during upload).
  3. Currently, only scans with max 442 368 000 Voxels can be analyzed. Use 512 x 512 resolution for reconstruction or a max. of 421 slices for 1024 x 1024 resolution.
  4. JAVA environment or Google Chrome are required for upload and analysis with LungTraX Platform.


Ideal Scan Parameters

Following these parameters will achieve the highest quality report possible

Highest Quality Report Parameters
PARAMETERSSIEMENSPHILIPSTOSHIBAGE
Kernel StandardB30BFC08Standard
Tube CurrentRegular Patient (<30 BMI): 80 mAs
Large Patient (>30 BMI): 100 mAs
*No tube current modulation
KV120
Slice Thickness0.625 mm
Reconstruction Interval
(Slice Spacing)
≤0.625 mm
PitchRange: 0.5–1.2Range: 0.5–1.2Range: 0.5–1.0Range: 0.5–1.375
Rotation or Gantry Speed (sec)≤0.5
Iterative ReconstructionNone
ContrastNone

Acceptable Scan Parameters

CT Scan Parameters WITHOUT Iterative Reconstruction

*No iterative reconstruction preferred

3–4mSv Dose
PARAMETERSSIEMENSPHILIPSTOSHIBAGE
Tube Current0 mA–900 mA
*No tube current
modulation preferred
20 mA–450 mA
*No tube current
modulation preferred
20 mA–600 mA
*No tube current
modulation preferred
30 mA–770 mA
*No tube current
modulation preferred
KV120
Dose ModulationCareDose ON
CarekV OFF
Z-Dom ONSURE Exposure ONSmart mA ON
PitchRange: 0.5–1.2Range: 0.5–1.2Range: 0.5–1.0Range: 0.5–1.375
Rotation or Gantry Speed (sec)≤0.5
Kernel Standard≤B45fB, C≤FC45Bone, Standard
Slice Thickness≤1.5 mm
*Thinnest slice possible preferred
Slice Spacing≤slice thickness
Average mSv<4.0
ContrastNone

How StratX Report Results Are Determined

The data presented in the StratX Report is based on Hounsfield units to measure fissure completeness, destruction scores, and lobar volumes. The quality of the uploaded CT images directly influences the Hounsfield unit of measurement. Below are examples of reasons why images will be rejected:

Motion Artifact Scan Rejection

Motion Artifact Image Distortion

Taking a CT scan while a patient is moving distorts the image quality, making it appear blurry or fuzzy. The image distortion directly impacts the accuracy of destruction, fissure, and volume scoring. A StratX Report will not be provided if a motion artifact is detected.

Inadequate Breath-Hold Scan Rejection

Expiratory Scans

Expiratory scans are noted by the raised diaphragm and lobar compression. Expiratory scans will lead to artificially low destruction scores and inaccurate lobar volumes. A StratX Report will not be provided if an expiratory scan is detected.

Raised Diaphragm

Contrast Scan Rejection

Scans with Contrast

If an image is taken with contrast, the vasculature will appear bright and will permeate into lung tissue resulting in
artificially low destruction scores. A StratX® Report will not be provided if a scan with contrast is detected.

Incorrect Kernel

Images reconstructed with an incorrect kernel may result in inaccurate destruction scores. In the example below, an incorrect kernel has added noise to the image, resulting in an artificially low destruction score. If an incorrect kernel is detected, a StratX Report will be provided with a warning that the report may be inaccurate.

How long does a StratX Report take?

A StratX Report is normally available within 3 working days. For a StratX Post-Treatment Report, please allow 10 working days.

LungTrax Platform Login

Log in to the LungTrax platform to access your StratX Reports.

LungTrax Support

Need help with your LungTrax account or a specific scan? Get in touch with us.
Please provide email address of the LungTraX Account, Scan ID or Patient ID for problems, with a certain scan, and a short description of the problem.

EMEA-EN-2325-v1